iM3 Catalogue - Europe - Flipbook - Page 122
Knowing when to replace your tips is essential, as tip wear will signi昀椀cantly reduce
scaler performance – as it alters the ultrasonic frequency.
ULTRA SCALING TIPS
Up to 1 mm:
Optimum
ef昀椀ciency
Each tip is designed for different cleaning needs, from coarse to 昀椀ne
depuration, for tackling both large deposits and interdental areas.
1-2 mm:
Time to
re-order
Over 2 mm:
Time to
dispose
CLICK HERE TO DOWNLOAD THE ULTRA SCALING TIP GUIDE
Perio - 1P
Universal - 1U
Universal - 2U
Universal - 3U
U801P
U801U
U802U
U803U
Suitable for the removal
of subgingival deposits.
Suitable for coarse depuration
of supragingival calculus.
Suitable for 昀椀ne depuration of
supragingival calculus.
To remove supragingival
calculus; suitable for
interdental cleaning.
Operating Settings:
Operating Settings:
Operating Settings:
Operating Settings:
Ideal: 10
Ideal: 30
Max: 30
Reorder new tips at www.im3vet.eu
Max: 40
Ideal: 20
Max: 30
Ideal: 20
Max: 30
! Please use the correct wrench supplied with the tip, and make sure it is fully tightened.
1
How to use the wrench to change tips
• Using the torque wrench, insert the scaler tip into the
handpiece coupling.
Rotate the torque wrench clockwise until it is tight; proceed
to rotate the wrench clockwise for 3 more turns. The tip is now
secured.
•
Do not ‘rock’ the torque wrench while installing the tip! Lateral
movement can result in a broken tip or damaged handpiece.
•
NOTE: Only iM3 scaler tips can be used with the Ultra Scaler.
CK
!
S C A L E R S & M I C R O M OTO R S
2
•
CLI
?
3
“I am seeing "Error 6" / my scaler light is flashing
- why is this?”
First check that your scaler tip is correctly attached
to the scaler coupling. If the issue persists, replace
the scaler tip with a new iM3 tip.
If this fails to solve your issue, please contact us at:
sales@im3vet.com
Page 122
www.im3vet.eu